Presentation is loading. Please wait.

Presentation is loading. Please wait.

Part No...., Module No....Lesson No L16.2: X線透視における防護の最適化

Similar presentations


Presentation on theme: "Part No...., Module No....Lesson No L16.2: X線透視における防護の最適化"— Presentation transcript:

1 Part No...., Module No....Lesson No L16.2: X線透視における防護の最適化
Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 (訳: 日本IVR学会 放射線防護委員会) 放射線診断・IVRにおける放射線防護 L16.2: X線透視における防護の最適化 Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

2 Part No...., Module No....Lesson No
はじめに Module title 主題 : X線透視における放射線防護 物理的および技術的要因の双方が患者と術者の被ばく線量に影響する. 適正な放射線防護の考え方と術者・スタッフの技量が、患者と術者の被ばくを低減するには欠かせない. Explanation or/and additional information Instructions for the lecturer/trainer 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

3 Part No...., Module No....Lesson No
Module title 内容 術者被ばくに影響する因子 患者被ばくに影響する因子 被ばく線量の実例 放射線防護具 放射線防護のルール Explanation or/and additional information Instructions for the lecturer/trainer 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

4 Part No...., Module No....Lesson No
Module title 概要 X線透視における実際的な放射線防護原則の適応に精通すること. Lecture notes: ( about 100 words) Instructions for the lecturer/trainer 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

5 Part 16.2: X線透視における放射線防護の 最適化
Part No...., Module No....Lesson No Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 Part 16.2: X線透視における放射線防護の 最適化 Topic 1: 術者被ばくに影響する因子 Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

6 おさらい: 吸収と散乱 X-Ray tube 患者に到達する1000の光子のうち,およそ100~200が散乱し,およそ20が検出器に到達し,残りは吸収される (= 被ばく線量) 散乱X線もまた逆二乗の法則に従うため,患者との距離をとることで安全性が増す 放射線学的に散乱線は主に管球側に向かう 16.2: 透視における防護の最適化

7 術者被ばくに影響する因子 (I) 透視室における術者への主な線源は患者である (散乱線). 散乱線は患者周囲で均一ではない.
患者周囲の線量率は多くの因子が複合的に関係する. 16.2: 透視における防護の最適化

8 術者被ばくに影響する因子 (II) 術者被ばくへの影響因子 術者の身長 患者との相対的位置関係 照射される患者の体積 X線管球の位置
kV, mA, 時間 (パルス透視の回数と波形) 吊下げ式遮蔽板やゴーグルの有効利用 16.2: 透視における防護の最適化

9 術者被ばくに影響する因子 (III) 角度の影響 散乱線の線量率は患者のX線入射部位の近くで高くなる 100 kV 1 mA
0.9 mGy/h 1 mA 0.6 mGyh 11x11 cm 0.3 mGy/h 患者からの距離 1m 患者の体厚 18 cm 16.2: 透視における防護の最適化

10 術者被ばくに影響する因子(IV) 照射野のサイズの影響 散乱線の線量率は照射野が広くなるほど増加する 100 kV 1 mA
11x11 cm 17x17 cm 17x17 cm 100 kV 0.8 mGy/h 1.3 mGy/h 1 mA 0.6 mGy/h 1.1 mGy/h 0.3 mGy/h 0.7 mGy/h 患者からの距離 1m 患者の体厚 18 cm 16.2: 透視における防護の最適化

11 術者被ばくに影響する因子(V) 散乱線の線量率は患者との距離が増せば低下する 距離による変化 100 kV 1 mA
mGy/h at 0.5m mGy/h at 1m 散乱線の線量率は患者との距離が増せば低下する 100 kV 1 mA 11x11 cm 16.2: 透視における防護の最適化

12 術者被ばくに影響する因子(VI) 一般に管球を患者寝台の下に置くことで術者の水晶体への高線量率を低減できる 最適な配置 検出器が上
X線管球が下 線量は1/3以下に 低下する 右と比較すると X線管球が上 検出器が下 一般に管球を患者寝台の下に置くことで術者の水晶体への高線量率を低減できる 16.2: 透視における防護の最適化

13 術者被ばくに影響する因子(VII) 一般に管球を患者寝台の下に置くことで術者の水晶体への高線量率を低減できる mGy/h mGy/h
X線管球 一般に管球を患者寝台の下に置くことで術者の水晶体への高線量率を低減できる mGy/h 100 kV 2.2 (100%) 1 m 2.0 (91%) 20x20 cm 1.3 (59%) mGy/h 1 Gy/h (17mGy/min) 1.2 (55%) 1.2 (55%) 1.2 (55%) 1 Gy/h 患者との距離 1m (17 mGy/min) 1.3 (59%) 20x20 cm 2.2 (100%) 100 kV 1 m 患者との距離 1m X線管球 16.2: 透視における防護の最適化

14 術者被ばくと患者被ばくは部分的に関連がある
決定因子 1. 使用するX線装置  装置の現状(保守)  装置の使用状況 2. 使用可能な防護具 3. 手技の回数と種類 4. 術者の技量と手技の プロトコル 術者被ばく 患者被ばく 16.2: Optimization of protection in fluoroscopy

15 術者被ばくと患者被ばくは部分的に関連がある
◆患者被曝は減少しても術者被ばくが増加することがある   (鉛手袋,可動遮蔽板などを   使わない場合) ◆一般に,患者被ばくが増加す れば,術者被ばくも増加する (多くの撮影枚数,長い透視 時間など). 術者被ばく 患者被ばく 16.2: Optimization of protection in fluoroscopy

16 術者被ばく・患者被ばくに影響する因子 (I)
それに伴って患者の皮膚線量と散乱線が増加する 患者の体格が増大すると 16.2: 透視における防護の最適化

17 術者被ばく・患者被ばくに影響する因子(II)
線量率が2倍もしくはそれ以上に増加する 通常透視モードから高線量率モードに変更すると 16.2: 透視における防護の最適化

18 術者被ばく・患者被ばくに影響する因子(III)
患者の入射部線量が2~6倍に増加 散乱線除去グリッドの使用 16.2: 透視における防護の最適化

19 Part No...., Module No....Lesson No Topic 2: 患者被ばくに影響する因子
Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 Part 16.2: X線透視における防護の最適化 Topic 2: 患者被ばくに影響する因子 Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

20 高ノイズモードから低ノイズモードへの変更
患者被ばくに影響する因子 ( I ) 高ノイズモードから低ノイズモードへの変更 (シネモードおよびDSA) 1画像につき2~10倍の線量増加 16.2: 透視における防護の最適化

21 従来型のアナログ透視からデジタル透視への変更
患者被ばくに影響する因子 (II) 従来型のアナログ透視からデジタル透視への変更 25%にまで線量率の低減が可能 16.2: 透視における防護の最適化

22 患者被ばくに影響する因子 (III) 12インチ (32 cm) 線量 100 9インチ (22 cm) 線量 150
患者の入射部線量の 相対値 検出器の口径 12インチ (32 cm) 線量 100 9インチ (22 cm) 線量 150 6インチ (16 cm) 線量 200 4.5インチ (11 cm) 線量 300 16.2: 透視における防護の最適化

23 検出器(蛍光増倍管)の視野径をより小さく変更
患者被ばくに影響する因子 (IV) 検出器(蛍光増倍管)の視野径をより小さく変更 患者の入射線量が3倍に増加する可能性 16.2: 透視における防護の最適化

24 Part No...., Module No....Lesson No Topic 3: 線量の実例
Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 Part 16.2: X線透視における防護の最適化 Topic 3: 線量の実例 Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

25 GE-CGR Advantix LCV での 1フレームあたりの線量
標準的な線量 4mGy/画像または 0.1mGy/フレーム Aモード: Bモード: Cモード: Dモード: 線量 1倍 線量 2.5倍 線量 5倍 線量 10倍 高ノイズ 低ノイズ 16.2: 透視における防護の最適化

26 透視における入射線量率の実例 低線量 10 mGy/分 中線量 20 mGy/分 高線量 40 mGy/分
GE-CGR Advantix LCV (透視) 低線量 10 mGy/分 中線量 20 mGy/分 高線量 40 mGy/分 16.2: 透視における防護の最適化

27 散乱線量率の実例 散乱線量は X線管球側で高い 16.2: 透視における防護の最適化

28 移動型Cアーム装置周囲の線量率の実例 検出器 1.2 等線量線の数値の単位は µGy/分 患者 3 6 12 X線管球
100 cm cm Scale 1.2 3 6 12 X線管球 等線量線の数値の単位は µGy/分 16.2: 透視における防護の最適化

29 Part No...., Module No....Lesson No Topic 4: 防護具
Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 Part 16.2:X線透視における防護の最適化 Topic 4: 防護具 Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

30 防護具(I) 甲状腺 プロテクタ 遮蔽板とゴーグル カーテン 16.2: 透視における防護の最適化

31 防護具(II) 100 kV 100 kV 透過線の強度 90 % 80 % 鉛入り手袋 同じ触感の場合 70 % 60 %
直接線 90 % 80 % 散乱線 鉛入り手袋 同じ触感の場合 100 kV 直接線 70 % 60 % 散乱線 タングステン入り手袋の タングステン 入り手袋 X線遮蔽力は鉛に比べ 約3倍良好!! 16.2: 透視における防護の最適化

32 複数個の個人線量計を装着することが推奨される
個人線量の測定 放射線 防護具 水晶体線量(オプション) 手指線量(適宜) エプロンの外・上部の頚部の第二線量計(適宜) 鉛エプロン下の個人線量計 複数個の個人線量計を装着することが推奨される 職業被ばくの線量限度 実効線量   5年間で平均 20 mSv/年 年等価線量限度  水晶体 150 mSv  皮膚 mSv  手足 mSv IVR手技による放射線障害の回避 ICRP 2000年 より 16.2: 透視における防護の最適化

33 Part No...., Module No....Lesson No Topic 5: 放射線防護のルール
Module title 国際原子力機関 診断およびIVRにおける放射線防護に関する修練教材 Part 16.2:X線透視における防護の最適化 Topic 5: 放射線防護のルール Part …: (Add part number and title) Module…: (Add module number and title) Lesson …: (Add session number and title) Learning objectives: Upon completion of this lesson, the students will be able to: . (Add a list of what the students are expected to learn or be able to do upon completion of the session) Activity: (Add the method used for presenting or conducting the lesson – lecture, demonstration, exercise, laboratory exercise, case study, simulation, etc.) Duration: (Add presentation time or duration of the session – hrs) Materials and equipment needed: (List materials and equipment needed to conduct the session, if appropriate) References: (List the references for the session) IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

34 X線透視室では, 吊下げ式遮蔽板,鉛エプロン, 手袋, 甲状腺プロテクタなどが
実践的放射線防護基準 (I) X線透視室では, 吊下げ式遮蔽板,鉛エプロン, 手袋, 甲状腺プロテクタなどが すぐに使えないといけない 起こりうる問題点: それらを適正に使用しなければならない 16.2: 透視における防護の最適化

35 実践的放射線防護基準 (II) 定期的な品質管理の体制を確立しなければならない
起こり得る問題点: スタッフはこれらの点検を定期化し,撮影室の十分な利用性を確保すべき 16.2: 透視における防護の最適化

36 各操作モードや検出器の入力面サイズにおける線量率を把握しなければならない 全ての操作モードにおいて適正な使用基準を確立しなければならない
実践的放射線防護基準 (III) 各操作モードや検出器の入力面サイズにおける線量率を把握しなければならない 全ての操作モードにおいて適正な使用基準を確立しなければならない 16.2: 透視における防護の最適化

37 実践的放射線防護規則 (IV) 重要な因子: X線管球と皮膚の距離 患者と検出器の距離 患者線量は以下の場合に増加する:
患者と検出器の距離が長い 16.2: 透視における防護の最適化

38 装置と専門医 (I) 装置の要因 専門医の要因 検出器の入力面の 各手技における 線量と画質 画像の記録数 技術サービスによる セッティング
16.2: 透視における防護の最適化

39 実際の検出器の性能によって線量率の増加が必要となる場合がある 実際の検出器の性能と必要な線量率を把握する
装置と専門医 (II) 装置の特性 専門医の役割 実際の検出器の性能によって線量率の増加が必要となる場合がある 実際の検出器の性能と必要な線量率を把握する 16.2: 透視における防護の最適化

40 装置と専門医 (III) 装置の特性 専門医の役割 自動輝度調整機能を適切に作動させること、場合によってその機能を停止できること
鉛手袋が照射野に入る場合など、高線量率となることを避けるため、その機能を適正に使用する 16.2: 透視における防護の最適化

41 装置と専門家 (IV) 装置の特性 専門医の役割 視野コリメーションの選択が容易なこと コリメーターの効果的な使用
16.2: 透視における防護の最適化

42 装置と専門医 (V) 装置の特性 専門医の役割 プロトコル  各手技における患者の全被ばく線量 グリッドによる影響 検出器の性能
ノイズレベル,パルス数/秒,パルス幅 など プロトコル  各手技における患者の全被ばく線量 16.2: 透視における防護の最適化

43 スタッフの被ばくリスク 装置の特性 専門医の役割 スタッフと患者の距離 および相対的位置関係 撮影室の形 遮閉板の厚さ X線装置の位置
16.2: 透視における防護の最適化

44 Part No...., Module No....Lesson No
Module title まとめ (I) X線透視装置を用いて手技を行う場合,X線束の方向,管球からの距離,検出器の視野径,透視装置の型式など,多くの物理的因子が患者および術者の被ばく線量に大きく影響する. 実際的な放射線防護のルールにより,これらの被ばくを軽減することができる. Let’s summarize the main subjects we did cover in this session. (List the main subjects covered and stress again the important features of the session) 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

45 Part No...., Module No....Lesson No
Module title まとめ (II): ”絶対的基準” 検出器を患者に近づける. 拡大モードを必要以上に使用しない. X線管球を患者からできるだけ遠ざける. 可能な部位では高電圧(kVp)撮像を用いる. 防護エプロンと個人線量計を装着すること,散乱線の最も強い場所を知っておくこと. 手技上可能な限り距離をおくこと. Let’s summarize the main subjects we did cover in this session. (List the main subjects covered and stress again the important features of the session) 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources

46 Part No...., Module No....Lesson No
Module title さらに詳しい情報を得るには Wagner LK and Archer BR. Minimising risks from fluoroscopic x rays. Third Edition. Partners in Radiation Management (R.M. Partnership). The Woodlands, TX USA 2000. Avoidance of radiation injuries from medical interventional procedures. ICRP Publication 85.Ann ICRP 2000;30 (2). Pergamon Radiation Dose Management for Fluoroscopically-Guided Interventional Medical Procedures, NCRP Report No. 168, National Council on Radiation Protection and Measurement. Bethesda, MD Interventional Fluoroscopy: Physics, Technology, Safety, S. Balter, Wiley-Liss, 2001 16.2: 透視における防護の最適化 IAEA Post Graduate Educational Course in Radiation Protection and Safe Use of Radiation Sources


Download ppt "Part No...., Module No....Lesson No L16.2: X線透視における防護の最適化"

Similar presentations


Ads by Google